The device displays a color and a text indication of the reliability of the IOP reading. If the deviation
between measurements is within normal limits, the color will be green and the text will read
'Deviation: OK'. If the deviation is slightly high, the reliability indication color will be yellow and the
text will read 'Deviation: DEVIATION'. If the deviation is high, the color will be red and the text will
read 'Deviation: REMEASURE'. The table below explains each reliability indication.
DISPLAYED TEXT DISPLAYED COLOR
Deviation: OK.
Green.
Deviation:
Yellow.
DEVIATION.
Deviation:
Red.
REMEASURE.
7.2 SETTINGS
Use the Settings menu to change the tonometer's settings.
To access the settings, go to Menu –> Settings and press the main button to
confirm.
Brightness - Change the display brightness
Increase or decrease the brightness using the up/down navigation buttons and
press the main button to confirm.
Volume - Turn the tonometer sounds on or off
Select on or off using the left/right navigation buttons and press the main button to
confirm. If the sound is off you will not hear the beep.
Date - Set the date
1. Select the month/day/year you want to change, by using the left/right navigation
buttons.
2. Change the month/day/year, by using the up/down navigation buttons.
3. Press the main button to confirm.
Time - Set the time
1. Select the hour/minute/second you want to change, by using the left/right
navigation buttons.
2. Change the hour/minute/second, by using the up/down navigation buttons.
3. Press the main button to confirm.
DEVIATION DESCRIPTION
< 15 % of
No or insignificant
the IOP.
deviation.
15-25 % of
Slight deviation; the
the IOP.
effect of the deviation is
unlikely to be relevant
to the result unless the
IOP is 19 or higher.
>25 % of the
The deviation is too
IOP.
high.
English
ACTION
-
If the IOP is 19 or
higher, take a new
measurement.
Take a new
measurement.