General Specifications
Display: 3-1/2 dgt LCD, 1999 counts
with function and unit indicators.
Zero Adjustment: automatic
Low Battery Indication:
battery voltage drops
below operating level.
Polarity indication: automatic,
"–" indicated; "+" implied
Display Update Rate, Digits: 2.5/sec,
nom.
Operating Temp.: 0 to 50°C, <80%
Relative Humidity
Storage Temp: -20°C to 60°C, <70%
RH, battery removed
Power Supply: four 1.5V AAA size
UM-4 R03 batteries
Battery Life (typical): 200 hrs zinc-
carbon.
Dimensions (HxWxD): 170 x 44 x 40
mm, (6.7" x 1.73" x 1.57")
Weight (including batteries):
250 grams (9 oz)
Housing material: ABS
Input connector: accepts standard
miniature thermocouple connectors
(flat blades, spaced 7.9mm, center
to center)
Accessories: one type K thermocou-
ple bead probe, soft carrying case,
operator' s manual
Specifications of supplied probe:
1.2m (4 foot) type K thermocouple
bead probe (teflon tape insulated).
SPECIFICATIONS
Maximum insulation temperature
260°C (500°F). Probe accuracy
±2.2°C or ±0.75% of reading
(whichever is greater) from 0°C to
800°C.
when
the European Community Directive:
89/336/EEC (Electromagnetic
Compatibility).
However, electrical noise or intense
electromagnetic fields in the vicinity
of the equipment may disturb the
measurement circuit. Measuring
instruments will also respond to
unwanted signals that may be pre-
sent within the measurement circuit.
Users should exercise care and take
appropriate precautions to avoid
misleading results when making
measurements in the presence of
electronic interference.
Electrical Specifications
Measurement range:
°C mode:
-50.0°C to 199.9°C; 0.1° resolution
-50°C to 1300°C; 1° resolution
°F mode:
-58.0°F to 199.9°F; 0.1° resolution
-50°F to 1999°F; 1° resolution
Resolution, selectable: 0.1° or 1°
Accuracy
atures over the range 18°C to 28°C /
64°F to 83°F, for 1 year, not including
– 10 –
EMC: This product com-
plies with requirements of
(specified for operating temper-