UNIT-MOUNTED MASTER FLUX
Connect the hose, exhaust gas circuit
mask to the hose in the center of the
cabinet.
Connect the circuit MF824ZMF to the
hose on mobile.
Connect the terminal to the suction.
SUCTION 
MF824ZMF
 
SETTING THE VALVE
Turn the ring perforated valve (SMFA236) until closing holes.
Start dispensing OXYGEN setting ~ 10 Nl / min.
Start the surgical aspirator, set the minimum intake.
Observe THE BAG ANESTHESIA.
Turn the ring perforated valve gradually opening the holes until you
get the ideal state of the BAG ANESTHESIA "HALF FULL HALF
EMPTY".
CAUTION:
By varying FLOW, it may be necessary, adjust the correct
intake flows of exhaled gases.
8
ACTIVE EXHAUST
WALL-MOUNTED MASTER FLUX
MF824ZMF connect the circuit to the discharge
circuit.
Connect the terminal to the suction.