en - Guide to the appliance
Reset operational
faults
Reset factory set-
tings
38

Control LEDs

The control LEDs (light emitting diodes) give status information about
the communication module.
LED
LIT UP GREEN
FLASHES GREEN The module is receiving or transmitting data.
LIT UP YELLOW
FLASHES YEL-
LOW
LIT UP RED
FLASHES RED
FLASHES altern-
ately
YELLOW and RED
or
GREEN and RED

Reset button

If the module experiences an operational fault during operation, the
reset button can be used to get it operational again.
To do so press the reset button for approx. 1-2 seconds.
 Do not press it for any longer than this as this would reset the
module to its factory default settings and you would then have to
reconfigure the interface.
The module will then restart (reset). This process can take up to
90 seconds. The LED next to the button will light up red during this
process.
If you cannot get the module operational again with the reset button
you will have to reset the module to its factory default settings.
 If the factory default settings are reset the interface will have to
be reconfigured.
Press and hold the reset button for approx. 5-6 seconds.
The LED will start to flash red. Resetting to default settings takes at
least 90 seconds.
Status
The module is ready for operation.
The module is receiving network data, the IP
address set is, however, not recognised. The
interface may need to be reconfigured.
Network connection being made.
Resetting in progress.
The module is initialising. This process can
take up to 90 seconds.
The module is in service mode and is not ready
for use. Switch the washer-disinfector off and
then back on again to come out of service
mode.