Reprocessing; Reprocessing Transmission Instruments With Spray Channels - Miele professional A 812 Instrucciones De Manejo

Tabla de contenido

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

  • MEXICANO, página 36

Reprocessing

transmission in-
struments with
spray channels
Reprocessing
For validation purposes please make sure that you follow the
loading instructions given on the template.
 To minimise the risk of injury, the mobile unit should be loaded
from the back forwards and unloaded in the reverse order.
 Any hollow items that have become disconnected from their in-
jector nozzles, adapters or holders during processing must be re-
processed.
To reprocess transmission instruments with spray channels, 1 to 4
A 815 nozzles with silicone hoses must be connected to the adapter.
 Ports that are not in use must be sealed with blind stoppers.
Check the silicone hoses for visible damage. Damaged silicone
hoses must be replaced prior to reprocessing.
Check that the transmission instrument and the spray channel are
free of blockages, following the specifications of the instrument
manufacturer.
Insert the transmission instrument into an A 803 holder or AUF 1
holder.
Connect the free end of the silicone hose to the lower end of the
spray channel.
en - Areas of application