Error Messages - ForaCare GD40 Manual Del Usuario

Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

Error Messages

MESSAGE
40
WHAT IT MEAN
Appears when the
batteries cannot provide
enough power for a test.
Appears when a used
test strip is inserted.
Problem in operation.
Appears when test strip
is removed while count-
ing down, or insufficient
blood volume.
Appears when the
wrong code strip is
inserted or other coding
errors.
Expired code strip.
Appears when ambient
temperature is below
system operation range.
Appears when ambient
temperature is above
system operation
range.
ACTION
Replace the batteries
immediately.
Repeat the test with a new test
strip.
Repeat the test with a new test
strip. If the meter still does not
work, please contact the cus-
tomer service for assistance.
Repeat the test with a new test
strip.
Make sure to insert the right
code strip for the meter.
Check if the code numbers on
the code strip and the strip vial
label are the same.
Repeat the test with a new
code strip.
System operation range is 50°F
to 104°F (10°C to 40°C).
Repeat the test after the meter
and test strip are in the above
temperature range.