Specifications (Cont.)
Noise Rejection*:
 Normal Mode Rejection Ratio > 60 dB at 50 Hz and 60 Hz when measuring DCV
 Common Mode Rejection Ratio > 60 dB from 0 Hz to 60 Hz when measuring ACV
 Common Mode Rejection Ratio > 100 dB at 0 Hz, 50 Hz and 60 Hz when measuring DCV
Operating Conditions:
 Temperature: 0 °C to 40 °C (32 °F to 104 °F)
 Relative Humidity (non-condensing): 80% maximum for temperatures up to 31 °C (88 °F),
  decreasing linearly to 50% maximum at 40 °C (104 °F)
 Altitude: 2000 m (6500') maximum
 Indoor use only
 Pollution Degree: 2
Storage Conditions:
 Temperature: –20 °C to 60 °C (–4 °F to 140 °F)
 Relative Humidity (non-condensing): 0% to 80%
 Remove battery.
Battery: Two 1.5 V batteries (AAA, NEDA 24A or IEC LR03)
Safety: Double insulation per IEC/UL EN61010-1 Ed. 3.0, IEC/EN61010-2-030 Ed. 1.0, IEC/
EN61010-2-033 Ed. 1.0, IEC/UL/EN61010-031 Ed. 1.1 and CAN/CSA-C22.2 No. 61010-1-12 Ed.
3.0 to Category II 1000V, CAT III 600V and CAT IV 300V AC & DC
Transient Protection: 6.5KV (1.2/50µs surge)
Overload Protections:
 μA and mA: 0.4A/1000V DC/AC rms, IR 30KA@ 1000V DC/AC rms; Dimensions: 6 x 32 mm
A: 11A/1000V DC/AC rms, IR 20KA@ 1000V DC/AC rms; Dimensions: 10 x 38mm
V & AutoCheck
: 1100V DC/AC rms
TM
mV, Ohm & others: 1000V DC/AC rms
Overvoltage Categories:
 Terminals (to COM) Measurement Category: V/mAµA / A : Category II 1000V, CAT III 600V and
CAT IV 300V AC & DC.
 E.M.C: Meets EN61326-1:2013 (EN55022, EN61000-3-2, EN61000-3-3, EN61000-4-2, EN61000-
  4-3, EN61000-4-4, EN61000-4-5, EN61000-4-6, EN61000-4-8, EN61000-4-11)
*Noise rejection is the ability to reject unwanted signals, or noise.
 • Normal mode voltages are AC signals that can cause inaccurate DC measurements. NMRR
 (Normal Mode Rejection Ratio) is a measure of the ability to filter out these signals.
 • Common mode voltages are signals present at the COM and + input terminals, with respect
 to ground, that can cause digit rattle or offset in voltage measurements. CMRR (Common
 Mode Rejection Ratio) is a measure of the ability to filter out these signals.