en - Intended use
Modules for reprocessing shafted instruments from the da Vinci®
surgical system can be used in the A 502 injector unit. The modules
are supplied with their own operating instructions.
15 screw attachments are provided at the rear of the injector unit for
further items used in robot-assisted surgery.
These are fitted with blind stoppers ex works. If required injector
nozzles, irrigation sleeves or silicone tubes can be connected in place
of the blind stoppers.
Internal cleaning and disinfection to an appropriate standard is only
possible with instruments which can be sufficiently dismantled and
with adequate use of injector nozzles, irrigation sleeves and adapters.
The reprocessing result of the inside of instruments must be checked
as necessary using a test other than a visual check.
The mobile unit can be used in the following washer-disinfectors:
– PG 8527
– PG 8528
Medical devices for reprocessing are referred to as the wash load if
they are not more closely defined.
Queries and technical problems
If you have any queries or technical problems that you would like to
discuss please contact Miele. Contact details can be found at the
back of the Operating Instructions for your washer-disinfector or go
to www.miele-professional.com.