5
The ear thermometer is a device capable
o f a c h i e v i n g i n f r a - r e d t e m p e r a t u r e
measurement when placed in the auditory
canal of a subject. It is a safe means of
measuring human body temperature through
the ear.
Specification
The standard we adopted is EN12470-
5:2003 Clinical thermometers -- Part 5:
Performance of infrared ear thermometers
(with maximum device).
This device complies with the requirements
of IEC 60601-1,IEC 60601-1-11 and IEC
60601-1-2.
Range of displayed temperature: 34.0° C ~ 44.0° C (93.2° F ~111.2° F)
Display L °C (L °F), when the temperature is under 34,0° C (93,2° F)
Display H °C (H °F) when the temperature is over 44,0°C (111,2°F)
Operating ambient:
Temperature: 16°C ~35°C (60.8°F ~95.0°F )
Humidity: ≤80% UR
Storage and transport condition:
Temperature: da -25°C a 55°C (da -13°F a 131°F )
Humidity: 85% UR
Display resolution temperature range: 0.1°C (0.1°F)
Accuracy: ±0.2°C (da 35.5°C a 42.0°C ) / ± 0.4°F (da 95.9°F a 107.6°F )
Display: liquid crystal display, 3,5 digits
temperature value: display the maximum temperature in measuring process
temperature unit: centigrade or fahrenheit
display of memory: last ten memories
low voltage warning: the LCD display
Power consumption: 9 milliwatt in measurement mode
Battery: one 3V Lithium battery (CR 2032)
Battery life: 2000 takes
Dimension: 110 mm x 35 mm x 35 mm
Net weight: Approx.37g
Beeper sign: on/off, measuring finish and low voltage warning, etc.
Self-testing sequence: Press the 'on/recall' button to turn on the thermometer and all of
symbol (See Fig A) should be displayed on the LCD in one second.
Manufacturing and Calibration date: See the label in the battery box.
Cautions
1. Before measuring, the thermometer shall be stabilized at least 30min under
operating condition.
2. Please keep your ear canal clear, if not, the measuring result shall be inaccurate.
3. The symbol '
Probe lens
On/recall button
Battery cover
and then a beep sound is heard
' on the LCD shows you can measure in the ear;