Operating Directions
Warning: Personal injury, Sealer damage, property damage and/or inaccurate test results may occur
unless Sealer is used according to these instructions.
• Turn Power Switch on. The amber Power Light should illuminate.
• Allow the Sealer to warm up and the green Ready Light to come on (up to 10 minutes). Sealer will not operate until both the amber
power light and the green Ready Light are illuminated, indicating that the unit has reached operating temperature.
• Place an empty Quanti-Tray or Quanti-Tray/2000 Rubber Insert on the Input Shelf with the large cutout facing away from the Sealer.
• Place a Quanti-Tray or Quanti-Tray/2000 filled with sample and DST* reagent onto the Rubber Insert, making sure that the Tray is
properly seated in the Rubber Insert, and with each well of the Tray in its corresponding Rubber Insert hole.
• Slide the Rubber Insert with Tray into the Sealer until the motor grabs the Rubber Insert and begins to draw it into the Sealer.
• In approximately 15 seconds, the Tray will be sealed and partially ejected from the rear of the Sealer. Remove the Rubber Insert and
Tray from the rear of the Sealer.
• If at any time you wish to reverse the motor drawing the Rubber Insert into the Sealer (for example, if a misaligned Tray is
accidentally fed into the Sealer), press and hold the Reverse Button. However, do not reverse the motor once the Rubber Insert
has been drawn fully into the Input Slot.
• Multiple Rubber Inserts can be run consecutively without pausing.
• Turn off Sealer when not in use.

Technical Specifications

Weight
Dimensions
35 lbs
12" H x 11" D x 16" W
16 kg
30cm H x 27cm D x 39cm W
Cleaning
The Rubber Insert may be autoclaved, or it may be cleaned with isopropyl alcohol or household bleach, taking the usual precautions
when handling such liquids. Be sure to clean the rubber insert if it shows any signs of built up grime.
Clean the outside of the Sealer with a soft, dry cloth. A soft cloth moistened with water, household bleach, or isopropyl alcohol may also be
used, taking the usual precautions when handling such liquids.
The Access Panel should only be opened by personnel qualified to clean the inside of the Sealer. Refer to the Preventive Maintenance
Instructions for directions on how to clean the inside of the Sealer. Cleaning should only be performed by trained per-
sonnel at your facility. Do not open the Access Panel or tilt the Sealer if sample is dripping from the unit.

Changing Fuses

Fuses are located in the fuse holder just above the Power Switch. If it is necessary to change a fuse, turn off the Sealer and
unplug the Power Cord from the Sealer before opening fuse holder. Use 6 Amp Buss MDL-6 fuses or equivalent in the 115V
unit and 4 Amp Buss GDC-4 fuses or equivalent in the 230V unit.
Ambient Temperature
32-90°F
0-32°C
Power
115V, 60Hz, 6Amp (model 99-10893-00)
230V, 50Hz, 3Amp (model 99-10896-00)
5 5 5 5 5