8
Information on EMC
8-1. EMC (Electromagnetic Compatibility) Information
Basic EMC
Phenomenon
standard or test
method
Mains terminal
disturbance
CISPR11:2015
voltage
Radiated
CISPR11:2015
disturbance
Harmonic Current
EN 61000-3-2:2014
Emission
IEC 61000-3-2:2014
Voltage change,
Voltage
EN 61000-3-3:2013
fluctuations and
IEC 61000-3-3:2013
Flicker Emission
Electrostatic
EN 61000-4-2:2009
Discharge
IEC 61000-4-2:2008
Immunity
EN 61000-4-
Radiated RF
3:2006+A2:2010
Electromagnetic
IEC 61000-4-
Field Immunity
3:2006+A2:2010
Immunity to
EN 61000-4-
Proximity Fields
3:2006+A2:2010
from RF wireless
IEC 61000-4-
Communications
3:2006+A2:2010
Equipment
Electrical Fast
EN 61000-4-4:2012
Transient/Burst
IEC 61000-4-4:2012
Immunity
EN 61000-4-5:2014
Surge Immunity
IEC 61000-4-5:2014
Operating
Port tested
mode
Operating
AC Mains
Operating
Enclosure
Operating
AC Mains
Operating
AC Mains
Operating
Enclosure
Stand by
Operating
Enclosure
Stand by
Operating
Enclosure
Stand by
Operating
AC Mains
Stand by
Operating
AC Mains
Stand by
Test level/
Test Voltage
requirement
120V , 60Hz
220V , 60Hz
Group1, Class A
230V , 50Hz
120V , 60Hz
220V , 60Hz
Group1, Class A
230V , 50Hz
230V , 50Hz
Class A
Pst: 1
Plt: 0.65
230V , 50Hz
Tmax: 0.5
dmax: 4%
dc: 3.3%
120V , 60Hz
± 8 kV/Contact
220V , 60Hz
± 2, ± 4, ± 8,
230V , 50Hz
± 15 kV/Air
120V , 60Hz
3 V/m
220V , 60Hz
80 MHz-2.7 GHz
230V, 50Hz
80% AM at 1 kHz
120V , 60Hz
Table 9 in
220V , 60Hz
IEC 60601-1-2:
230V , 50Hz
2014
120V , 60Hz
± 2 kV, 100 kHz
220V , 60Hz
repetition
230V , 50Hz
frequency
Line to Line
± 0.5 kV, ± 1 kV
120V , 60Hz
Line to Ground
220V , 60Hz
± 0.5 kV, ± 1 kV, ±
230V , 50Hz
2
kV
31