PROBLEM
Start / Stop
"10s"
blinking
Start / Stop
3 beeps
Malfunctions during charging phase.
blinking
Hand piece placed in the charging
unit but the charging phase does
not start.
POSSIBLE CAUSE
The LED pack could be defective.
Overheating protection.
3 beep
Overheating protection.
Defect
REMEDY
Verify the LED pack functionality by
means of the following procedure:
press the "Start / Stop" button conse-
cutively for 5 times
"10s"
"10s"
Start / Stop
Note:
No manual reset
Stop
Alignment key.
GB