Probe Installation - Hach POLYMETRON Serie Manual Del Usuario

Tabla de contenido

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

  • MEXICANO, página 47
Note: For illustration purposes, cable part number 08319=A=00xx is shown in
1 External shielding
2 Internal shielding
3 Internal electrode
The cable is available in lengths of 5, 10 or 20 meters and must be connected in compliance with the
following table:
Function
External shielding
Internal shielding
Internal electrode
External electrode
Note: Refer to the user manual delivered with the transmitter for a detailed description of the cable connectors on
the transmitter.

5.3 Probe installation

In
Figure 5
on page 8,
C indicate:
• A: Ideal installation - perfect immersion of the electrode surfaces.
• B: Good installation - satisfactory immersion of the electrode surfaces.
• C: Poor installation - incomplete immersion of the electrodes, the conductivity will be too low.
5.3.1 On piping
Immerse the internal electrode completely in the process sample. For a 90° installation, take into
account the dimensions (see
Note: In the following illustrations, the arrows indicate the sample flow direction.
Installation example for the 8315 probe
Pt100
Pt100
Figure 6
on page 8 and
Dimensions
on page 5).
4 External electrode
5 Pt100
6 Pt100
White (orange tip)
Figure 7
on page 9 the annotations A, B and
Figure
4.
Color
White (red tip)
White (yellow tip)
Red
Black
Blue
English 7

Publicidad

Tabla de contenido
loading

Tabla de contenido