The Dräger X-zone 5000 is equipped with a pump as
an option (refer to "Order list" on page 49.)
Commissioning and performing the
measurement
If the device and pump is being used in hazardous
zones, requiring devices complying with "Equipment
Protection Level" (EPL) Ga (corresponds to 1G Zone 0,
for example), only the Viton hose (order no. 12 03 150)
should be used with a maximum length of 30 m.
Pump operation only with filter (order no. 83 19 359)!
Otherwise there is a risk of damage to the pump.
If no filter is used during pump operation, the warranty
on the pump becomes invalid.
z Release the lock on the holder.
z Remove the holder.
z Press the
key of the Dräger X-zone 5000 and
OK
hold it for approx. 3 seconds.
– The Dräger X-zone 5000 is in STANDBY mode.
z Place the Dräger X-am 5000 into the device
receptacle.
z Place the holder (pump) onto the housing.
z Fasten the lock at the holder (pump).
z Switch on the Dräger X-am 5000 in accordance
with the Instructions for Use.
– If the device is in the ON mode, the pump will be
automatically switched on via a switching contact
on the holder (pump).
– Pump LED flashes red/green.
z Connect the sampling
hose with the water
trap for the hose set
(order no. 68 05 473)
and screw the
connecting bush onto
the filter in a clockwise
direction.
z A pump test will then need to be performed.
The pump test will start automatically.
NOTICE
The pump test must be performed within 60 seconds
otherwise a device alarm is issued.
38
WARNING
CAUTION
Pump test
NOTICE
Pump test is performed in the same way as for the
Dräger pump X-am 1/2/5000. During the pump test, in
the default setting the acoustic alarm is automatically
reduced to 80 dB (A).
– The pump LED flashes red/green,
accompanied by a signal tone.
z Seal or kink the intake flange or
sampling hose for approx.
2 seconds.
– The pump LED lights red.
– Unblock the intake flange/
sampling hose.
If the test was successful:
– The pump LED lights green,
accompanied by an acoustic
acknowledgment signal.
If the test was not successful:
– The pump LED lights red,
accompanied by a continuous
tone.
– The pump switches off
automatically.
Ending operation
z Switch off the Dräger X-am 5000 in accordance with
the Instructions for Use.
– The visual and the audible alarm will be activated for
a short time before the device switches off.
– The Dräger X-zone 5000 switches to STANDBY
mode.
z Unscrew the sampling hose or the Dräger probe
from the filter.
z Remove the holder (pump) by releasing the lock.
– The pump switches off.
z Place the holder onto the housing.
z Fasten the lock at the holder.