Guidance and Manufacturer`s declaration electromagnetic emissions

The Laser systems „3B LASER PEN 200 mW" laser system is intended for use in the electromagnetic environment speci-
fied below. The customer or the user of the "3B LASER PEN 200 mW" laser system should assure that it is used in such an
environment.
Emissions test
RF emissions
CISPR 11
RF emissions CISPR 11
Harmonic emissions
IEC 61000-3-2 (
*
)
Voltage fluctuations/flicker emissions
*
IEC 61000-3-3 (
)
*
(
) Note: The requirement is only valid for equipment with power input between 75 and 1000 W

Guidance and manufacturer's declaration-electromagnetic immunity

The Laser system „3B LASER PEN 200 mW" is intended for use in the electromagnetic environment specified below. The
customer or the user of the "3B LASER PEN 200 mW" should assure that it is used in such an environment.
Immunity test
Electrostatic discharge
(ESD)
IEC 61000-4-2
Electrical fast transient/
burst
IEC 61000- 4-4
Surge
IEC 61000-4-5
Voltage dips, short inter-
ruptions and voltage vari-
ations on power supply
input lines
IEC 61000-4-11
Power frequency
(50/60 Hz) magnetic field
IEC 61000-4-8
Note: Uτ is the a.c. mains voltage prior to application of the test level.
70 70
Compliance
Group 1
Class B
Not applicable
(< 75 W)
Not applicable
(< 75 W)
IEC 60601 test level
± 6 kV contact
± 8 kV air
± 2 kV for power supply
lines
± 1 kV for input/output
lines (not applicable, no
input/output lines)
± 1 kV line(s) to line(s)
± 2 kV line(s) to earth (not
applicable, no earth con-
nection)
< 5 % Uτ for ½ cycle
(> 95 % dip in Uτ)
40 % Uτ for 5 cycles
(60 % dip in Uτ)
70 % Uτ for 25 cycles
(30 % dip in Uτ)
< 5 % Uτ for 5s
(> 95 % dip in Uτ)
3 A/m
Electromagnetic environment-
guidance
The "3B LASER PEN 200 mW" laser sys-
tem uses RF energy only for is internal
function. Therefore, its RF emissions
are very low and are not likely to cause
any interference in nearby electronic
equipment.
The „3B LASER PEN 200 mW" laser sys-
tem is suitable for use in all establish-
ments, including domestic establish-
ments and those directly connected to
the public low-voltage power supply
network that supplies buildings used
for domestic purposes.
Compliance level
Electromagnetic
environment- guidance
± 6 kV contact
Floors should be wood, concrete
± 8 kV air
or ceramic tile. If floors are covered
with synthetic material, the relative
humidity should be at least 30 %.
± 2 kV for power supply
Mains power quality should be that
lines
of a typical commercial or hospital
± 1 kV for input/output
environment.
lines (not applicable, no
input/output lines)
± 1 kV line(s) to line(s)
Mains power quality should be that
± 2 kV line(s) to earth
of a typical commercial or hospital
(not applicable, no earth
environment.
connection)
< 5 % Uτ for ½ cycle
Mains power quality should be that
(> 95 % dip in Uτ)
of a typical commercial or hospital
40 % Uτ for 5 cycles
environment. If the user of the „3B
(60 % dip in Uτ)
LASER PEN 200 mW" laser system
70 % Uτ for 25 cycles
requires continued operation dur-
(30 % dip in Uτ)
ing power mains interruption, it is
recommended that the „3B LASER
< 5 % Uτ for 5s
PEN 200 mW" laser system be pow-
(> 95 % dip in Uτ)
ered from an uninterruptible power
supply or a battery.
3 A/m
Power frequency magnetic fields
should be at levels characteristics of
a typical location in a typical com-
mercial or hospital environment.