RESOLUTION GUIDE:
Possible anomalies in the functioning
INCIDENCE
The device does
not turn on.
The stimulation
is low or non-
existent.
18
Manual Electroestimulador ET-250.indb 18
CAUSE
The product does not have
batteries or they are inserted
incorrectly.
The batteries have died.
The on/off dial remains
"OFF".
Damaged equipment.
Low intensity level.
On the screen there is a "P".
Badly connected electrodes.
Dirty electrodes.
Electrodes in poor condi-
tions.
Short-circutied electrodes (in
contact with one another).
The distance between the
electrodes is excessive.
The batteries are about to
run out.
SOLUTION
Insert 4 AAA batteries
with the correct pole.
Replace the batteries
with new ones
Turn the dial clockwise
to turn on the device.
Contact the Daga tech-
nical service.
Slowly increase the in-
tensity through the dial
of the corresponding
channel.
Select a specific type of
treatment (6) or a dedi-
cated programme (4).
Make sure that the
cable is correctly con-
nected to the appliance
and to the electrodes.
Clean the electrodes
with cold water as
shown in this manual.
Replace the electrodes
with new ones. You
can acquire them and
the normal point of
purchase or contact the
DAGA customer service
centre.
Position these electro-
des leaving a greater
distance between them.
Position these elec-
trodes leaving a lower
distance between them.
Replace the batteries
with new ones