6. Press the Start button.
The LOW CURRENT message
will turn up:
7.
Turning the Test current
control clockwise, increase
current until obtaining the
desired value or until the
current bargraph reads 100
divisions. The bargraph will
show the test current value
as the percentage of the
nominal value selected by
using the Range selector.
8.
The equipment should be connected to the mains supply in order
to work with 100 A.
9.
The lowest current for measuring is 10% of the range nominal
value. It is important to consider that the measurement errors
increase while test current decreases. The error is specified for
test currents higher than 80%.
10. The Display will show the resistance value measured and the
unit used (Ω [ohms], mΩ [milli-ohms] or µΩ [micro-ohms]).
11. The value can be retained in
the second line of the display
by pressing the button Hold.
If you press this button
again, the value will be
erased.