Ocultar thumbs Ver también para Polar Care Wave:
Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

CLEANING

WARNING:
Do not apply cold-compression pad to patient during the cleaning procedure. Exposure
to cleaning agent may cause injury.
CAUTION: Do not apply a direct stream of any liquid onto the unit's lid, submerge the unit, or allow any
liquid to pool on the top of the lid. The unit's lid contains all of the unit's electronics.
Clean as necessary or if soiled or debris is present in the unit. If using well water, more frequent cleanings
may be required.
Polar Care Wave – Exterior Surfaces
(container, lid, tubing, cold-compression pad, and y-adapter)
Cleaning agents
for home setting:
Cleaning agents
for clinical setting:
Procedure:
Polar Care Wave – Internal Components
(pump system, internal tubing, inside of cold-compression pad and y-adapter)
Cleaning agents:
Procedure:
• Mild soap
• Water
• Use only mild soap and water on cold-compression pad
• PDI sani cloth wipes
• Quaternary ammonium
• Follow the manufacturer's instructions and precautions for the cleaning
agent you select.
• Use selected cleaner and a soft cloth to wipe surfaces of container, lid,
tubing, connectors, and cold-compression pad.
• Wipe dry.
• Bleach
• Water
• Connect cold-compression pad to unit.
• Fill the unit with cold tap water to the ice line and add 1/2 teaspoon of
bleach.
• Do not add ice.
• Run the cold therapy for at least 10 minutes.
• Roll the cold-compression pad while still connected and squeeze the water
back into the container.
• Disconnect the cold-compression pad, pour the water out of the unit, and
wipe dry.
15

Publicidad

Tabla de contenido
loading

Tabla de contenido