PRODUCT SPECIFICATIONS
Electrical Specifications (Also see Electromagnetic Immunity Test Table on page 21)
Unit Rated: 12VDC, 1A
Power supply is specified as part of ME Equipment
Manufacturer: HDP
Back up supply: FRIWO
Rated Input: 100-240VAC, 50-60Hz
Standards Compliance:
Electrical Classifications:

Environmental Requirements

Transport & Storage:

Operating:

Performance Specifications

Expected Service Life:
Average operating temperature of a
Polar Care Wave cold-compression pad:
Note: The Foot/Ankle pad at the cold setting may operate at temperatures above 60°F which only
provides a cooling sensation.
Average peak compression pressure of
the cold-compression pad:
20
Model: HDP12-MD12010C
Model: FW8000M/12
Designed to conform to applicable requirements of: EN 60601-1,
Edition 3.1 and 60601-1-2, 4th Edition
Class II, Continuous Operation. Not suitable with flammable
anesthetics. Group 1, Class B.
–25°C to 5°C, and
5°C to 35°C at a relative humidity of 10% to 90%, non-condensing;
>35°C to 70°C at a water vapor pressure up to 50 hPa
5°C to 40°C (41°F to 104°F )15-90% RH non-condensing,
700 hPa to 1060 hPa
Indoor use only up to 2000m altitude
There is no warm up time required from when the Wave is stored at
the minimum storage temperature and when the product is used.
1 year from date of first use (unit and accessories).
≥45°F at the colder setting
≥50°F at the cold setting
50 mmHg at the regular setting
25 mmHg at the low setting
Output: 12VDC, max. 1.0A
Output: 12VDC, max. 1.0A