Medical Electrical Equipment needs special precautions regarding EMC and needs to be installed and put into service according the EMC information
provided in the following.
Portable and mobile RF communication equipment (e.g. cell phones) can affect Medical Electrical Equipment.
The use of accessories and cables other than those specified (other than boso original parts) may result in increased emissions or decreased immunity of
the unit.
The boso unit is intended for use in the electromagnetic environment specified below. The customer of the user of the boso unit should
assure that it is used in such an environment.
Emissions test
RF emissions
CISPR 11
RF emissions
CISPR 11
Harmonic emissions
IEC 61000-3-2
Voltage fluctuations/flicker
emissions IEC 61000-3-3
Recommended separation distances between portable and mobile RF communications equipment and the boso unit
The boso unit is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. The customer or the user of
the boso unit can help prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF communications
equipment (transmitters) and the boso unit as recommended below, according the the maximum output power of the communications equipment.
Rated maximum output
power of transmitter
W
0,01
0,1
1
10
100
For transmitter rated at a maximum output power not listed above, the recommended separation distance d in metres (m) can be estimated using the
equation applicable to the frequency of the transmitter, where p is the maximum output power rating of the transmitter in watts (W) according the the
transmitter manufacturer.
NOTE 1 At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects
and people.
Guidance and manufacturer's declaration – electromagnetic emissions
Compliance
Group 1
Class B
Class A
Complies
Separation distance according to frequency of transmitter
150 kHz to 80 MHz
d = 1,2 P
0,12
0,38
1,2
3,8
12
Electromagnetic environment – guidance
The boso unit uses RF energy only for its internal function. There-fore, its RF
emissions are very low and are not likely to cause any interference in nearby
electronic equipment.
The boso unit is suitable for use in all establishments, including
domestic establishments and those directly connected to the public low-
voltage power supply network that supplies buildings used for
domestic purposes.
m
80 MHz to 800 MHz
d = 1,2 P
0,12
0,38
1,2
3,8
12
800 MHz to 2,5 GHz
d = 2,3 P
0,23
0,73
2,3
7,3
23
33