Icare ic100 tonometer manual
ACCESSORIES
SKU
PRODUCT DESCRIPTION
104
Box of 100 probes
7217
TA011/TV011 collar
7218
TA011/TV011 collar narrow
540
Probe base
559
Wrist strap with lock
525
Aluminium case, Icare ic100
7169
Battery cover & screw, Icare ic100
623
IOP pad, Icare ic100
543
Probe base cleaning container
565
Silicone grip - white, Icare ic100
548
Screw driver, Icare
577
USB memory stick
544
Probe base cover
TECHNICAL INFORMATION
Type: TA011
Dimensions: 24 - 29 mm (W) * 35 - 95 mm (H) *
215 mm (L).
Weight: 140 g (without batteries), 230 g (4 x AA
batteries).
Power supply: 4 x AA non-rechargeable
batteries, 1.5V alkaline LR6.
Measurement range: 7-50 mmHg
Accuracy: ±1.2 mmHg (≤20 mmHg) and ±2.2
mmHg (>20 mmHg).
Repeatability (coefficient of variation): <8 %.
Accuracy of display: 1 mmHg.
Display unit: Millimeter of mercury (mmHg).
The serial number is on the inside of the battery
compartment cover.
There are no electrical connections from the
tonometer to the patient.
The device has BF-type electric shock
protection.
Operation environment:
Temperature: +10 °C to +35 °C
Relative humidity: 30 % to 90 %
Atmospheric pressure: 800 hPa-1,060 hPa
Storage environment:
Temperature: -10 °C to +55 °C
Relative humidity: 10 % to 95 %
Atmospheric pressure: 700 hPa-1,060 hPa
Transport environment:
Temperature: -40 °C to +70 °C
Relative humidity: 10 % to 95 %
8
WEIGHT
89 g
4 g
1 g
4 g
4 g
800 g
6 g
38 g
3 g
26 g
15 g
44 g
1 g
Atmospheric pressure: 500 hPa-1,060 hPa
Mode of operation: continuous
PERFORMANCE DATA
The performance data is obtained from
a clinical study, performed according to
American National Standard ANSI Z80.10-
2003 and International Standard ISO 8612.2
for tonometers. The study was performed in
the Department of Ophthalmology, Helsinki
University Central Hospital. In the study, 158
patients were measured. The mean paired
difference and standard deviation (Goldmann-
Icare) were -0.4 mmHg and 3.4 mmHg. A
scattergram and Bland-Altman plot of the
results is shown in Figure 20.
MAINTENANCE
Follow local regulations and recycling
instructions regarding the disposal or recycling
of the Icare tonometer and accessories.
WARNING
The tonometer should only be opened by
qualified service personnel. It contains no
userserviceable parts, apart from the batteries
and a probe base. The Icare tonometer requires
no routine servicing or calibration other than
changing the batteries at least every 12 months
or changing or cleaning the probe base. If
servicing is necessary, contact qualified service
personnel or your local Icare representative.
DIMENSIONS
53 x 109 x 36 mm
18 x 18 mm
17 x 18 mm
7 x 38 mm
10 x 10 x 270 mm
240 x 280 x 72 mm
110 x 25 x 12 mm
50 x 53 x 16 mm
20 x 56 mm
45 x 35 x 113 mm
16 x 90 mm
98 x 11 x 93 mm
19 x 11 mm