Technical Characteristics - TUCKY 21 Instrucciones De Uso

Tabla de contenido

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

technical characteristics

Product type
Reference
Medical device
Normal condition of use
Storage/transport conditions
Atmospheric pressure
Rate of humidity
Battery
Range of temperature displayed
Display resolution
Accuracy
Technologie Bluetooth
Compatibility
USB Cable
Charge voltage
Protection index
Lifetime of Tucky
Lifetime of ADH-21 adhesives
Software classification
ESSAIS ELECTROMAGNETIC EMISSION TESTING
RF Emissions (CISPR 11)
RF Emissions (CISPR 11)
Harmonic current emissions
Voltage fluctuation, flicker emissions
IMMUNITY TESTING
Electrostatic discharge (ESD) CEI 61000-4-2
RF radiated IEC 60601-4-3
Magnetic field at the network frequency IEC 60601-4-8
Warranty : please refer to the general sales terms, available on the website www.e-takescare.com
Not intended to be sterilised. Subject to modifications without notice.
Medical thermometer in direct mode
Tucky-21
Classe IIa
Ambient operating temperature: +15°C to 35°C
Technical accuracy cannot be guaranteed if the device is used outside the temperature and
humidity ranges indicated
Tucky-21W: Temperature: -15°C to 55°C ± 2°C. Humidity: 15% to 95% ± 5°C RH
ADH-21: Temperature: +10°C to 27°C ± 2°C. Humidity: 40% to 60% ± 5°C RH
Can operate at an atmospheric pressure which does not exceed 1 (760 à 1060 hPa)
Relative humidity range of 15% to 85% (without condensation)
Device with internal power supply: lithium battery. Capacity: 15mAh
+24°C to +44°C
0.1 °C
Accuracy for the range of temperatures displayed (maximum error in the laboratory): ±0.2°C
from 32°C to 42 °C then +/- 0.3°C outside this range
4.0
Compatibility: from iOS 7.3 and Android 4.3
USB 2.0 Cable Type A Male to Type B Male
5V
IP22: Protected against solid objects larger than 12 mm.
Protected against vertically falling drops of water up to 15°C
500 charging cycles
2 years
Class B
COMPLIANCE
ENVIRONMENT
Group 1
The RF emissions are very low and are not likely
to cause interference in nearby electronic equipment.
Class B
Compliant
Not applicable
Battery power
Not applicable
Battery power
TESTING LEVEL ACCORDING TO CEI 60601
± 8 kV at contact
± 2,4,8 and 15 kV in the air
10V/m de 80 MHz à 2,5 GHz
30 A/m
LEVEL OF COMPLIANCE
Compliant
Compliant
Compliant