6. Specifications
Model
Display
Measurement range
Accuracy
Inflation
Deflation
Measurement method
IP classification
Power source
Battery life
Operating temperature / humidity
Storage temperature / humidity / air pressure
Weight
Dimensions
Cuff circumference
Memory
Contents
Applied part
Protection against electric shock
Notes:
• These specifications are subject to change without notice.
• In the clinical validation study, the 5th phase was used on 85 subjects for determination of diastolic blood
pressure.
• This device has not been validated for use in pregnancy.
BP765
HEM-7311-ZSA
LCD digital display
Pressure: 0 to 299 mmHg
Pulse: 40 to 180 beats / min.
Pressure: ±3 mmHg or 2% of reading
Pulse: ± 5% of display reading
Fuzzy-logic controlled by electric pump
Automatic pressure release valve
Oscillometric method
IP 20
4 "AA" batteries 1.5V or AC adapter (INPUT AC100-240V 50/60Hz 0.12A)
Approximately 1000 measurements (using new alkaline batteries)
50°F to 104°F (10°C to 40°C) / 15 to 90% RH
-4°F to 140°F (-20°C to 60°C) / 10 to 95% RH / 700 to 1060 hPa
Monitor : Approximately 1 lbs 6 5/8 oz. (640 g) not including batteries
Arm cuff : Approximately 6 oz. (170 g)
Monitor : Approximately 7 1/4" (w) × 3 7/8" (h) × 9" (l)
(183 mm × 99 mm × 230 mm)
Arm cuff : Approximately 5 3/4" × 23 3/8" (air tube: 29 1/2")
(145 mm × 594 mm (air tube: 750 mm))
9" to 17" (220 to 420 mm)
Up to 100 per user
Monitor, arm cuff, AC adapter, instruction manual, quick start guide
= Type BF
Internally powered ME equipment (When using only the batteries)
= Class II ME equipment (AC adapter)
EN
27