39
Anhang
Hinweise zur elektromagnetischen
Verträglichkeit (EMV)
Guidance and manufacturer's declaration – electromagnetic immunity –
XENON NOVA
®
175 model 20 1315 20 is intended for use in the electromagnetic environment specified below.
The customer or user of the XENON NOVA
EN/IEC 60601
Immunity test
test level
3 V rms
Conducted RF
IEC 61000-4-6
150 kHz to < 80 MHz
3 V/m
Radiated RF
IEC 61000-4-3
80 MHz to 2.5 GHz
Note: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection by structures, objects
and people.
a Field strengths from fixed transmitters, such as base stations for radio [cellular/cordless] telephones and land mobile radios, amateur radio, AM
and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic environment due to fixed
RF transmitters, an electromagnetic site survey should be considered. If the measured field strength in the location in which the unit is used exceeds
the applicable RF compliance level above, the unit should be observed to verify normal operation. If abnormal performance is observed, additional
measures may be necessary, such as reorienting or relocating the unit.
b Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m.
Appendix
Electromagnetic Compatibility
(EMC) Information
Table 204
for equipment and systems that are not life-supporting
®
175 model 20 1315 20 should ensure that it is used in such an environment.
Compliance level
Portable and mobile RF communications equipment should be used
no closer to any part of the XENON NOVA
includ ing cables, than the recommended separation distance calcu-
lated from the equation applicable to the frequency of the transmitter.
Recommended separation distance:
d
3 V rms
d
3 V/m
d
Where P is the maximum output power rating of the transmitter in
watts [W] according to the transmitter manufacturer and d is the
recommended separation in meters [m].
Field strengths from fixed RF transmitters, as determined by an
electromagnetic site survey
level in each frequency range
Interference may occur in the vicinity of equipment marked with
the following symbol:
Anexo
Indicaciones sobre compatibilidad
elec tromagnética (CEM)
Electromagnetic environment - guidance
®
175 model 20 1315 20,
[
]
=
3
5 .
3 /
P
150 kHz to < 80 MHz
[
]
=
3
5 .
3 /
P
80 MHz to < 800 MHz
[ ]
=
7
3 /
P
800 MHz to 2.5 GHz
a
, should be less than the compliance
b
.