Operation
• Fitting the battery
Change the batteries with this display
indication. Check position of the O-Ring when
opening/ closing the battery chamber lid!
Switch the unit on using the ON/OFF key.
The display shows: "0" . Make one Zero for all the
methods:
Swirl the instrument several times under the
water surface allowing the sample chamber
to completely fill. Take the instrument out of the
water.
Close the blue lid tightly.
Press the ZERO/TEST key. "000" flashes for
approx. 8 seconds. The display shows "0.0.0" .
Select the test required using the MODE key:
Cl (free Chlorine/ total Chlorine) → pH → TA →
CyA → Cl ... (Scroll).
Open the blue lid and add the appropriate tablet
straight from the foil.
Method
: Free chlorine
Cl
Cl
: Total chlorine
pH : pH Value
CyA : Stabilizer
TA : Alkalinity-M
{ Crush the tablet using a clean stirring rod.
Stir until the tablet is dissolved.
| Close the lid and swirl the instrument gently
for approx. 15 seconds.
} Afterwards keep the instrument steady and wait
for 15 seconds when testing ClF/pH/CyA and
2 minutes when testing Clt and TA.
Press the "Zero/Test"-key. The "---" symbol flashes for
approx. 6 seconds.
4
(2 x AAA (1.5 V), LR03)
appropriate tablet
DPD No.1
DPD No.1 + DPD No.3
Phenol Red
CyA-Test
Alka-M
The result appears in the display.
Result below the lowest limit of the
LOW
measuring range
HIGH
Measuring range exceeded
Repeat Test
Press the "Zero/Test"-key
Change Method
Press the "Mode"-key to scroll down.
New zero:
Press the "Zero/Test"-key for 2 sec.
• Maintenance
1. Keep the instrument in a clean condition.
2. No solvent, abrasive materials or brushes
should be used to clean the instrument. Clean
the sample chamber using a soft tissue.
3. Dry the instrument when not in use, store
under ambient environmental conditions.
4. Remove battery if instrument is being stored
for an extended period.
• Trouble shooting
1. Use only blister tablets with black printing.
2. Do not touch the tablets. This can cause
contamination and give pour results.
3. Close the blue lid on top of the sample chamber
tightly before zeroing/reading.
4. Always ensure that the sample chamber, the lid
and the stirring rod are thoroughly rinsed when
changing from one test to the other.
Warranty
The waterproof Electronic Pooltester is warranted to
be free from manufacturing defects for two years.
Conversion Table
mmol/l K
1 mg/l CaCO
0.02
3
°dh
°e
S4.3
0.056 0.07
0.10
°f
5