Metrix MTX 1032-B Manual De Instrucciones página 14

Sonda diferencial
Tabla de contenido

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

Description of MTX 1032-C (continued)
Description
The MTX 1032-C probe has 2 differential channels.
A LED indicates that it is switched on.
The probe does not include its own display facilities; it is designed for
connection to the CH1 and CH2 inputs of the oscilloscope so that the
differential signal present on the +/- inputs can be viewed.
Furthermore, this probe has two switches for selecting 1/10 or 1/100 as
the attenuation coefficient.
For each differential channel, the front panel of the MTX 1032-C probe
includes:
The MTX 1032-C is a stand-alone unit powered by the 230 V
50 Hz / 60 Hz mains supply.
1. Power up the probe using the mains lead. Check that the
Functional
description
2. Connect the probe's output "I" coaxial cable to input CH1 on the
3. Connect the probe's output "2" coaxial cable to input CH2 on the
4. Connect the red BNC lead to the positive input of the probe and the
5. Capture the signal to be measured.
6. Select the appropriate attenuation coefficient for the signal to be
7. When the measurements are finished, disconnect the leads. Switch off
1. Set the attenuation coefficient of the MTX 1032-C probe to 1/10.
Compensation of
1/10 probes
2. Connect the first 1/10 probe [ref. 3-(1)] to the positive input of the
Under-compensated probe
3. Compensate the probe and then disconnect it from the Probe Adjust
4. Connect the second 1/10 [ref. 3-(2)] to the negative input, capture
5. Connect the two 1/10 probes simultaneously to the Probe Adjust
Compensated probe
Over-compensated probe
To use the 1/10 probes of the MTX oscilloscope with the MTX 1032-C
probe, remove the sprung hook and fit the crocodile clamp.
14
2 BNC connectors for the +/- inputs
1 BNC for output to the oscilloscope
1 attenuation coefficient selection switch: 1/10, 1/100
corresponding LED comes on.
oscilloscope.
oscilloscope.
black BNC lead to the negative input of the probe.
measured (1/10 or 1/100) using the probe switch.
The signal is then displayed on the oscilloscope screen.
the probe and then disconnect the oscilloscope.
MTX 1032-C probe and then capture the Probe Adjust signal.
output.
the Probe Adjust signal and compensate the probe.
output to check that the adjustment is balanced: the amplitude of the
square signal is cancelled out.
English
AC
MTX 1032-B - MTX 1032-C

Publicidad

Tabla de contenido
loading

Este manual también es adecuado para:

Mtx 1032-c

Tabla de contenido