Manual Lubrication
• Remove rotary instruments from instrument chuck before cleaning, disinfection and lubrication.
• Place the instrument in a cloth or paper towel so as to catch any debris and foreign matter expelled by the atomizer jet.
• Insert the nozzle of the lubricant into the back of the instrument sleeve.
• Spray the STATCARE Spray for 1-2 seconds.
• Run the handpiece for 15-20 seconds to remove any residual debris and oil.
Manual Care of the Chuck:
It is recommended that the chuck system be cleaned and lubricated at least once a week. After removing the rotary
instrument, use the tip of the nozzle of the STATCARE spray can to spray into the opening of the chuck. Thereafter, please
follow the care and maintenance instructions provided by the manufacturer.