Recommended reprocessing methods according to DIN EN ISO 17664
The operator should be aware of personal health risks when reprocessing
contaminated dental handpieces and should wear gloves and eye
protection.
• Wipe off any debris or blood contamination at chair side.
• Do not immerse dental handpieces in disinfection solutions.
• Reprocess immediately after use.
• Remove any rotary instruments (burs, files, etc.) before reprocessing.
Cleaning and Disinfection
Please note that STATIS dental handpieces should not be cleaned in an
ultrasonic bath.
Attention: Do not use products containing acetone, chlorine or bleach as
disinfectants
*ALPRO WL-clean, ALPRO WL-cid and ALPRO WL-dry are registered trademarks of ALPRO MEDICAL GMBH, Mooswiesenstrasse 9, 78112 St. Georgen, Germany.
Manual External Cleaning and Disinfection
Cleaning:
After each use, the device must be thoroughly scrubbed either under water
(potable water) or with 60 - 70% alcohol solution.
Disinfection
Do not use products containing acetone, chlorine or bleach as
disinfectants. When applying disinfection sprays/wipes, follow the
instructions for use of these materials.
FOR GERMAN USERS:
In order to efficiently reprocess dental handpieces, internal cleaning and
disinfection can be achieved through the use of ALPRO WL-clean*, ALPRO
WL-cid* and ALPRO WL-dry*.