Guidance and manufacturer's declaration – electromagnetic emissions
The PRT2000 equipment is intended for use in the electromagnetic environment specifi ed
The customer or the user of the PRT2000 should assure that it is used in such an environment.
Emissions Test
Compliance
RF Emissions
Group 1
CISPR 11
RF Emissions
Class B
CISPR 11
Harmonic emissions
Not Applicable
IEC 61000-3-2
Voltage fl uctuations /
Not Applicable
fl icker emissions
Non-Life Support Equipment Separation Distance Calculation
(3Vrms / 3V/m compliance)
Separation distance according to frequency of transmitter (m)
150 kHz to 80 MHz
Rated maximum
in ISM bands
output power of
transmitter (W)
3.5
d = [
V
1
0.01
/
0.1
/
1
/
10
/
100
/
8
below.
Electromagnetic environment – guidance
The ME equipment uses RF energy only for its
internal function. Therefore, its RF emissions are
very low and are not likely to cause any interfer-
ence in nearby electronic equipment.
Complies
The ME equipment is
solely battery powered.
80 MHz to 800 MHz
800 MHz to 2.5 GHz
3.5
d = [
] P
d = [
] P
E
1
0.12
0.38
1.2
3.8
12
Guidance and manufacturer's declaration – electromagnetic immunity
The PRT2000 is intended for use in the electromagnetic environment specifi ed below.
The customer or the user of the PRT2000 should assure that it is used in such an environment.
Immunity test
Electrostatic dis-
charge (ESD)
IEC 61000-4-2
Radiated RF
IEC 61000-4-3
Conducted RF
IEC 61000-4-6
Electrical fast
transient
7
] P
IEC 61000-4-4
E
1
Surge
0.23
IEC 61000-4-5
0.73
2.3
Power frequency
magnetic fi eld
7.3
IEC 61000-4-8
23
Voltage dips, short
interruptions and
voltage variations on
power supply input
lines
IEC 61000-4-11
IEC 60601 test
Compliance
level
level
±6kV Contact
Complies
±8kV Air
3 V/m 80MHz to
Complies
2.5GHz
Not Applicable
3Vrms 150kHz to
(no electrical
80MHz
cabling)
±2kV power line
Not Applicable
±1kV I/O lines
±1kV
diff erential
Not Applicable
±2kV common
3 A/m
Complies
>95% dip 0.5 cycle
60% dip 5 cycles
Not Applicable
70% dip 25 cycles
95% dip 5 sec.
Electromagnetic environ-
ment – guidance
Floors should be wood,
concrete, or ceramic tile.
If fl oors are covered with
synthetic material, the
relative humidity should be
at least 30%.
Field strengths outside
the shielded location
from fi xed RF transmitters,
as determined by an
electromagnetic site survey,
should be less than 3 V/m.
Interference may occur in
the vicinity of equipment
marked with the following
symbol:
Separation distance
calculation provided above.
If a known transmitter is
present the specifi c distance
can be calculated using the
equations.
The ME equipment
is solely battery powered.
Power frequency
magnetic fi elds should be at
levels characteristic of a
typical location in a typical
commercial or hospital
environment.
The ME equipment
is solely battery
powered.
9