Stylus does not register
icons on the display
Icons missing from the
display
Calibration procedure
failed or cannot be
completed
Touch screen
display requires
calibration
Icon has been
de-selected
Turbine may be
faulty
Turbine not
fitted tightly
to calibration
syringe
Calibration
syringe does
not have an
inspiratory seal
or seal is leaking.
Shaft of the
syringe being
pushed down
86
Select the calibration
check icon and choose
touch screen, and follow
the instructions.
Hold stylus on to the
blue area of the display,
a list will appear, ensure
required icon is selected.
Repeat calibration
procedure, if problem
still persists, replace
turbine or return unit for
servicing
Ensure the syringe is
fitted to the turbine
using the adapter
supplied part # 861427.
(supplied with the
syringe)
Ensure you are using a
Vyaire syringe.
The syringe should be
emptied and filled with
one smooth stroke,
avoid pushing down on
the shaft or banging
at the end of each
manoeuver.