Specifications and Accuracy (cont'd)
Continuity Test: 0 to 50 kΩ
Test Current (typical): 1.0 µA
Open Circuit Voltage: Less than 1.0 VDC
Indication: Tone and LED
Operating Conditions: Tester must be operated within these limits for
accurate indication.
Temperature: –10 °C to 55 °C (14 °F to 131 °F)
Relative Humidity: 20% to 96%
Altitude: 2000 m (6500') maximum
Pollution Degree: 2
Indoor use only
Storage Conditions: –10 °C to 55 °C (14 °F to 131 °F),
less than 96% relative humidity
Remove battery.
Battery: Two 1.5 V batteries (AAA, NEDA 24A or IEC LR03)
Measurement Categories: CAT IV, 600 V / CAT III, 1000 V
CE Compliance:
EMC: EN61326 (1997, 1998/A1), EN61000-4-2 (1995), and EN61000-4-
3 (1996). Performance is specified in an RF field of 3 V/m.
Performance above 3 V/m is not specified. For GT-95e LCD display
only, add 45 digits to specified accuracy in an RF field of 3 V/m.
Safety: EN61243-3 Voltage Class B, EN61010-1 and IEC 61010-1 2nd
Edition (2001)
Type of Protection: IP51
GT-95e Only
RCD Check (for 230 V nominal circuits only—this feature is disabled
above 270 V and below 120 V):
Load Current: 30 mA nominal at 230 V
Impedance: 7.7 kΩ nominal, protected by PTC
Internal Load: 6.9 W @ 230 V
Note: This feature provides a convenient method to check RCD circuit
breakers. It is not intended to verify breaker effectiveness.
Effectiveness of RCD circuit breakers should be checked by RCD
testing, measuring, or monitoring equipment under the scope
of IEC 61557-1 and IEC 61557-6.
Display: 3-digit LCD (999 maximum count)
Display Turn-on Voltage: 8 VAC, 4.5 VDC
Voltage Resolution: 1 V (AC or DC)
LCD Accuracy:
ACV: ± (2.5% of reading + 4 V)
DCV: ± (1% of reading + 2 V)
14