7 Troubleshooting

Problem / Message
Printout unreadable
Pilot lamp does not light
Pilot lamp flashes
Timeout
--------------
---------------
Over Load
------------
-------------
Under Load
------------
-----------
Transmission Error
--
--
GlobalTestSupply
www.
Possible cause
Correction
 Ribbon worn or
 Change ribbon cartridge
entangled.
or tense the ribbon
properly.
 Service life of
printing unit at an
 Please contact OHAUS
end.
service.
 No power.
 Switch printer on.
 Power supply fault.
 Please contact OHAUS
service.
 No connection
 Make sure that the
printer device.
printer and the
instrument is connected.
 Printer did not find
 Switch the printer off/on.
the settings of the
 Alternatively change the
connected
settings of the printer or
instrument.
the connected
instrument. (see chapter
2.4)
 Data transfer has
 Move balance to more
not completed in
stable location.
the last 45 s.
 Check the balance.
 Balance in
 Check the sample
overload range.
weight.
 Zeroing out of the
 Change the zero range
zero range.
of the balance.
 Change the setting of
the printer from zero to
tare.
 Balance in
 Check the balance.
underload range.
Weighing pan is
not in position.
 Instrument does
 Check interface
not understand
parameter settings on
printer instruction.
printer and instrument.
EN-15