Find Quality Products Online at:
SF40A Printer
7 Troubleshooting
Problem / Message
Printout unreadable
Pilot lamp does not light
Pilot lamp flashes
Timeout
--------------
---------------
Over Load
------------
-------------
Under Load
------------
-----------
Transmission Error
--
--
GlobalTestSupply
www.
Possible cause
Correction
Ribbon worn or
Change ribbon cartridge
entangled.
or tense the ribbon
properly.
Service life of
printing unit at an
Please contact OHAUS
end.
service.
No power.
Switch printer on.
Power supply fault.
Please contact OHAUS
service.
No connection
Make sure that the
printer device.
printer and the
instrument is connected.
Printer did not find
Switch the printer off/on.
the settings of the
Alternatively change the
connected
settings of the printer or
instrument.
the connected
instrument. (see chapter
2.4)
Data transfer has
Move balance to more
not completed in
stable location.
the last 45 s.
Check the balance.
Balance in
Check the sample
overload range.
weight.
Zeroing out of the
Change the zero range
zero range.
of the balance.
Change the setting of
the printer from zero to
tare.
Balance in
Check the balance.
underload range.
Weighing pan is
not in position.
Instrument does
Check interface
not understand
parameter settings on
printer instruction.
printer and instrument.
EN-15
.com