EN-16
Problem / Message
Not Executable
--------
------
Date/Time Not Set
----
---
Memory Error
---------
---------
Memory full
---------
----------
Load/Unload Weight
--
---
Unit change not
allowed!
Out of range
---------
---------
GlobalTestSupply
www.
SF40A Printer
Possible cause
Correction
 Instrument cannot
 Check the balance.
execute printer
instruction.
 Printer error
 Check the instrument.
messages.
 Reset the printer.
 Set date/time
 Back-up battery
depleted
Please contact OHAUS
service.
 Statistics or
 Start a new Statistics or
Totaling Memory is
Totaling application.
full.
 No sample or the
 Put a new sample or
same sample is on
remove it and put it
the weighing pan.
again on the pan.
 Changed unit
 Reset the balance.
during Statistics or
 Change the unit back to
Totaling
the first one.
application.
 Sample weight is
 Sample weight must be
out of range
in the range of 70% -
130% of the current
average value.
.com