EyeSys Mini-IMG gauge dimension
FRG-700 gauge dimension
J5 – IN-OUT
This connector carries all the input and output sig-
nals to remote control the TPS-compact.
It is a 15-pins D type connector; the available sig-
nals are detailed in the table, the following para-
graphs describe the signal characteristics and use.
PIN N.
1
START/STOP (+)
2
START/STOP (-)
3
INTERLOCK (+)
4
INTERLOCK (-)
5
SPEED SETTING (+)
6
SPEED SETTING (-)
7
SOFT START (+)
8
SOFT START (-)
9
+24 Vdc
10
NORMAL OUTPUT (relay)
V
= 125V
max
11
PROGRAMMABLE SET POINT
12
NORMAL OUTPUT
13
FAULT OUTPUT
14
PROGRAMMABLE ANALOG
SIGNAL (+)
15
•
GROUND
•
PROGRAMMABLE ANALOG
SIGNAL (-)
Signal description
START/STOP: input signal to start or stop the
pump. Without the supplied cover connector the
turbomolecular pump automatically starts, if you
want to stop the turbo-pump from outside, you
have to connect the START/STOP (+) signal to the
+24 Vdc pin and the START/STOP (-) signal to the
GROUND pin (this is the default configuration of
the cover connector).
INTERLOCK: input signal to control the pump ro-
tation. With the supplied cover connector the IN-
TERLOCK (+) signal is connected to the +24 Vdc
pin and the INTERLOCK (-) signal to the
GROUND pin; in this condition the pump is
stopped.
SOFT START: this input is used to provide a "soft
start" to the pump; in this condition the ramp-up
time could be up to 20 min.
SPEED SETTING: PWM input signal to set the
pump speed. The PWM signal characteristics must
be the following:
• frequency: 100 Hz +/-20%
64/74
TECHNICAL INFORMATION
SIGNAL NAME
OUTPUT
I
= 30 mA
max
I
= 200 mA
max
87-901-010-01
INPUT/
IN
IN
IN
IN
IN
IN
IN
IN
OUT
OUT
OUT
OUT
OUT
OUT
OUT