Rocking Shaker & 3D Shaker
automatically halted and an alert will sound.
If these operations above are normal, the
instrument is ready to operate. If not, the
instrument
may
transportation, please contact technical support of
manufacturer/supplier.
Note:
The speed value can be altered during
shaking.
6. Faults
•
When switched ON, the instrument does not work.
Check whether the power cable is connected.
Check whether the fuse is broken.
•
Shaking function suddenly stop.
LCD displays "Er 03", indicate the current
failure is "overload protection".
Pressing the speed control knob or timer control
knob first to close shaking function, and then
restart shaking function after lowered load. If
be
damaged
during
shaking function cannot start, then repeat the
process and gradually reduce load.
7. Maintenance and Cleaning
•
Proper maintenance can keep instrument working
properly and lengthen its lifetime.
•
Do not spray cleanser into the instrument when
cleaning.
•
Do not remove the power line when cleaning. Only
use recommended cleansers:
Dyes
Construction
materials
Cosmetics
Foodstuffs
17
Isopropyl alcohol
Water containing
tenside / Isopropyl
alcohol
Water containing
tenside / Isopropyl
alcohol
Water containing
tenside