3.5/ AUTO OFF

The PAC 11 has an Auto Off feature which switches off 10 minutes after
the clamp has been switched on, without the controls being used.
When the clamp is switched off by this automatic function, the switch ’
must first be set to the OFF position before being switched on again.
This function can be overridden by the user when switching on. Simply
press the auto zero button  at the same time as setting the switch ’ from
the OFF position to the 1 mV/A or 10 mV/A position. If the green indicator
‘ flashes whilst the zero reset button is being pressed, this indicates that
the auto off function has been inhibited.

4/ SPECIFICATIONS

Ranges
input/output
ratio
40 A
10 mV/A
400 A
1 mV/A
4.1/ RÉFERENCE CONDITIONS
- Temperature: 18...28°C
- Humidity rate: 20...75% RH
- Battery voltage: 9 V ± 0.1 V
- Position of conductor: centred on the markings of the clamp
- Magnetic field: Earth's DC field
- Absence of external AC magnetic field
- Absence of electric field
- Measurement for a DC current or an AC sinusoidal current ≤ 65 Hz
- Impedance of the measurement instrument : ≥ 1 MΩ and ≤ 100 pF

4.2/ OPERATING CONDITIONS

The instrument must be used in the following conditions to satisfy the
safety of the user and the metrological performance:
- Use indoors
- Working altitude:≤ 2000 m
- Transportation altitude:≤ 12 000 m
- Environmental conditions: see graph below
Storage
% RH
Stockage
90
80
70
60
50
40
30
20
10
0
-50
-40
-30
-20
-10
measurement extent
AC rms
peak max
0.2...40 A
0.2...60 A
0.5...400 A
0.5...600 A
Use
Utilisation
0
10
20
30
40
50
60
Temperature
température en °C
12
DC
0.4...60 A
0.5...600 A
Reference
Réference
°C
70
80
90

Publicidad

Tabla de contenido
loading

Este manual también es adecuado para:

Pa c 11

Tabla de contenido