This device has been validated to withstand 5,000 cleaning and
disinfection cycles using the recommended disinfecting wipe/towelette.
The tested number of cycles is estimated by 2 disinfection cycles per day
for 5 years the expect life of device.
Stop using the meter if you see any signs of deterioration. For example:
• meter can not be turned on,
• LCD display cracks or becomes cloudy,
• buttons no longer function,
• meter outer casing cracks,
• data can not be transmitted to PC,
• color or paint/printing on housing is abnormal, or
• scratches or abrasions on meter are higher than acceptable.
Please contact the customer service for a replacement meter if any of the
signs of deterioration are noticed.
NOTE :
• Do NOT clean and disinfect the meter while performing tests.
• If the meter is being operated by a second person, the meter and lancing
device should be decontaminated prior to use by the second person.
• Do NOT allow cleaning and disinfecting solution to get in the test slot.
• If you do get moisture in the test strip slot, wipe it away with a corner of tissue.
• Always dry the meter thoroughly before using it.
• Do not spray the meter directly with cleaning solutions especially those
containing water (i.e. soapy water), as this could cause the solution to enter
the case inside and damage the electronic components or circuitry.