9
Technical data
Power consumption
Rated voltage / fuse type
Nominal frequency
Protective class
Pollution degree
Pressure hose connection
Nominal pressure
Max. operating pressure
Operating temperature of water
Height
Width
Depth
Weight
10
Information on malfunctions
Error messages
Error
Cause
E 99
Breakage / short circuit of sensor
or
Automatic rapid deactivation (software) owing to
excess temperature.
E 01
Process start attempted without timer.
E 02
Process start attempted without (previous)
heating function.
Polymerization unit for hot- and cold-curing resins
650 W
100 V, 2 x T8A / 250 V
120 V, 2 x T6.3A / 250 V
230 / 240 V, 2 x T4A / 250 V
50 / 60 Hz
1
2
3 to 10 bars
2 bars
2.8 bars
55°C (131°F) / 90°C (194°F) / 100°C (212°F)
deviation +/– 2.5°C (4.5°F) or approx. 5%
Dimensions
approx. 340 mm
approx. 360 mm
approx. 350 mm
approx. 6 kg
Remedy
•
Switch off the unit.
•
Check water filling level (see chapter 7.2 Filling with water).
•
Allow the unit to cool down approx. 10 min.
•
If the error can not be eliminated, inform your responsible service agent
(see chapter 11 Service).
•
Set process timer (see chapter 7 Working with the Palamat elite).
•
Activate standby heating function (see chapter 7.3.3 Starting a process).
•
If the error can not be eliminated, inform your responsible service agent
(see chapter 11 Service)
Operating Instructions
Required space
approx. 500 mm
approx. 370 mm
approx. 500 mm
GB
- 11 -