Chapter 2,
Operation
1. Verify that tubing is connected and power is supplied.
2. Verify that automatic regulator is ON.
The calibration cylinder concentrations must match
the instrument calibration setting; otherwise, instrument
readings may be incorrect.
3. Insert instrument into the test station.
NOTE: The following test sequence occurs:
Test Sequence
Alarm Test
•
Audible alarm and visual alarm functions are checked
•
ALARM TEST LED lights
•
LED flashes amber during test
•
Green LED indicates passing
•
Red LED indicates failure
•
Solid amber LED indicates that Altair 4 alarms
are turned OFF.
Gas Test
•
Sensor response to gases is verified
•
GAS TEST LED lights
•
LED flashes amber during test
•
Green LED indicates passing
•
Red LED indicates failure.
Overall Pass/Fail Test Indication
•
PASS OR FAIL LED lights
•
Green LED indicates passing
•
Red LED indicates failure
•
Checkmark symbol appears on the instrument display
when it is successfully bump checked
" WARNING
2-1