Assembly / Commissioning

Tubing installation with joint
type of sensor
setting ?
Access to all setting menus can be blocked with an access code !
meas. parameter
setting ?
32
Fig. 4
8
Commissioning
8.1

Presettings

Check first that the sensor is correctly connected to the
DULCOMETER
type of sensor :
ICT 1
sensor data
are changed!
Check that the right sensor type has been selected (see "type of
sensor setting?")
mr: meas. range
cc: cellconstant
α: temperature-
coefficient
Set the measurement range of the sensor (see "meas. parameter
setting?". Bypass the second parameter set (press Enter))
IMPORTANT
Check settings in all menus when changing the range allocation.
®
D1C (do not open the sensor).
20 mS/cm
mr :
cc :
1.120/cm
α :
2.10% °C
15 mS/cm 23.8 °C
mr 2 : 200 µS/cm
cc 2 : 1.120 /cm
2.10% /°C
α:
1106 µS/cm 25.0 °C