Drying
Manual Drying
Apply dry air to the internal and external surfaces until surface water is no longer visible.
Lubrication
After cleaning and / or disinfecting the the AM20 it is important to maintain/lubricate with SciCan Statcare Spray or use an automated care system
such as the Statmatic maintenance unit.
Manual Lubrication
• Place the motor in a cloth or paper towel so as to catch any debris and foreign matter expelled by the atomizer jet.
• Insert the nozzle of the lubricant into the back of the instrument sleeve.
• Spray the STATCARE Spray for 1-2 seconds.
• Run the handpiece for 15-20 seconds to remove any residual debris and oil.
Sterilisation of the STATIS AM20
Please note that the STATIS AM20 motor CANNOT be sterilized!