8. Technical specifications
Regulation
Output range
Programming range
Output resolution
Programming resolution
Accuracy
Line regulation
Load regulation
Ripple
Short term stability
Long term stability
Start current check
Sudden load change check
Ground leakage check
Output protection
Recovery after power failure
Ambient operating temperature
Ambient operating humidity
Ambient operating pressure
Mains requirements
Power consumption
Dimensions (WxDxH)
Weight
Maximum voltage, current and power with automatic cross-over at preset limits
Voltage:
0-600 V DC,
Current:
0-400 mA
Power: 0-100 W
Voltage:
6-600 V DC
Current:
1-400 mA
Power:
1-100 W
Time:
00:01-500 h
Volthour:
1-500 000 Vh
Voltage:
1 V
Current:
100 µA, 0-39.9 mA
1 mA, 40-400 mA
Power:
1W
Voltage:
1 V
Current:
1 mA
Power:
1 W
Time:
1 min, 00:01-99:59,
1h, 100-500 h
Volthour:
1 Vh, 1-9999 Vh,
100 Vh, 10.0-99.9 kVh
1 kVh, 100-500 kV
Voltage:
1%, ± 3 V
Current:
1%, ± 1 mA
Power:
2%, ± 1 W
Timer:
0.1% ± 1 min, 00:01-99:59 h
0.1%, ± 10 min, 100-500 h
< 0.2%
< 1% at load change 10-90% of maximum load
< 1% at 600 V
< 0.2% /10 h after warm up
< 1% /year
Resistance not greater than 0.2 MΩ at 40 V
(current less than 200 µA). Can be disabled
Leakage not greater than 500 µA
Fully protected against any overload conditions
Duration < 8 s, The program continues automatically
Duration ≥ 8 s, The program continues after manual restart
4-40 °C
0-95%
68-106 kPa, maximum altitude of 2000 m
100-120/220-240 V ~ ; 50/60 Hz
Max 140 W
250 x 315 x 95 mm
3.8 kg