FRx Trainer
10-04-07
15:12
18
behavior
Trainer says ANALYZING
before pads are applied
Trainer does not say
ANALYZING when pads are
applied to the
manikin adapter.
Trainer's voice instructions
are distorted.
Battery depletes
prematurely.
FRx TRAINER 861307
Page 18
To troubleshoot the FRx Trainer
The following table provides troubleshooting tips.
possible cause
• User touched the adhesive gel on
the pads and activated the sensing
to manikin.
circuitry of the Trainer.
• Pad wires are twisted.
• Pad wires are not separated.
• Pads are touching each other.
• Pads placement is incorrect.
• Pads are old, damaged, or use
discontinued blue gel.
• Manikin adapter is broken.
• In scenarios 3 and 8, pad has not
been removed and repositioned.
• Pad wires are twisted.
• One of the batteries is installed
backwards.
• Trainer is stored so that On/Off
button is inadvertently pressed.
• One of the batteries is installed
backwards.
recommended action
• Avoid touching the adhesive gel
on the pads.
• Untwist the pad wires.
• Separate the pad wires between
the pads and the red label.
• Separate pads.
• Check for proper pad and manikin
adapter placement. Remove and
reposition pads or adapter
if indicated.
• Replace training pads.
• Replace manikin adapter.
• Remove and reapply one pad.
• Untwist the pad wires.
• Reinstall batteries according to
the diagram in the battery holder.
• Store the Trainer so as to prevent
inadvertent pressure on the
On/Off button.
• Reinstall batteries according to
the diagram in the battery holder.