Accuracy Specifications - Coviden Nellcor DS100A Instrucciones De Uso

Tabla de contenido

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

  • MEXICANO, página 19
2. Failure to apply the DS-100A properly may cause incorrect measurements.
3. Using the DS-100A in the presence of bright lights may result in inaccurate measurements. In
such cases, cover the sensor site with an opaque material.
4. Reusable sensors must be moved to a new site at least every 4 hours. Because individual skin
condition affects the ability of the skin to tolerate sensor placement, it may be necessary to
change the sensor site more frequently with some patients. If skin integrity changes, move
the sensor to another site.
5. Intravascular dyes or externally applied coloring such as nail polish, dye, or pigmented
cream, may lead to inaccurate measurements.
6. The performance of the DS-100A is compromised by motion.
7. Do not apply tape to secure the sensor in place or to tape it shut; venous pulsations may
lead to inaccurate saturation measurements.
8. As with all medical equipment, carefully route cables to reduce the possibility of patient
entanglement or strangulation.
9. Do not use the DS-100A or other oximetry sensors during MRI scanning. Conducted current
may cause burns. Also, the DS-100A may affect the MRI image, and the MRI unit may affect
the accuracy of oximetry measurements.
10. Do not alter or modify the DS-100A. Alterations or modifications may affect performance or
accuracy.
If you have questions regarding any of this information, contact Covidien's Technical Services
Department, or your local Covidien representative.
Warranty
To obtain information about a warranty, if any, for this product, contact Covidien's Technical
Services Department, or your local Covidien representative.

Accuracy Specifications

For the accuracy specification range when used with Nellcor™ monitors, refer to information
provided with the monitor, or (in the U.S.), contact Covidien's Technical Services Department.
Outside the U.S., contact your local Covidien representative.
For the accuracy specification range when used with a monitor other than those manufactured
by Covidien, consult the information provided by the manufacturer of that instrument.
Note: Refer to the instrument operator's manual for complete instructions for use of the sensor
with that monitor.
5

Publicidad

Tabla de contenido
loading

Este manual también es adecuado para:

Nellcor ds100a-1

Tabla de contenido