Ambient Conditions During Operation
Temperature
Ambient pressure
Relative humidity
Power Supply
Mains power connection
Maximum power consumption
Dimensions
Device (width x depth)
Height (with hood closed)
Height adjustment range
Weight
Mattress area
Adjustment range mattress
Inclination of the mattress tray
Side wall height
X-ray tray
Display
Operating Modes
Air temperature regulation
Radiant warmer
Skin temperature regulation
Skin temperature monitoring
Operating Mode
Heated Mattress (Optional)
Humidifier regulation
Value
20 to 35 °C (68 to 95 °F)
620 to 1,100 hPa (9.0 to 16.0 psi)
20 to 95 %, non-condensing
100 V to 240 V, 50/60 Hz
1000 W
1,154 mm x 690 mm (45.43 in x 27.17 in)
1,850 to 2,250 mm (72.83 to 88.58 in)
400 mm (15.75 in)
Electrical height adjustment for optimum adaptation to clinical
users
<140 kg (308.6 lbs) (without options and accessories)
450 mm x 690 mm (17.72 in x 27.17 in)
90 mm (3.54 in)
Continuously variable electrical mattress tray tilt mechanism (up to
13°) to both sides and bed height adjustment
205 mm (8.07 in) with mattress tray in bottom position
464 mm x 379 mm (18.27 in x 14.92 in)
For larger film formats remove the x-ray tray and insert detector.
Tray is accessible from outside for taking x-rays without disturbing
the patient.
264 mm (10.4 in) Color- Touchscreen
Measured Value Display
Setting range 20 to 39 °C (68 to 102.2 °F)
<28 °C (82.4 °F), after confirmation
>37 °C (98.6 °F), after confirmation
Measurement range 13 to 45 °C
(55.4 to 113 °F)
with 2 heating elements for an optimal and even heat distribution
on the mattress area, automatic adaption of radiant warmer when
mattress is tilted
Setting range Off, 10 % to 100 %
Pre-warm mode:
100 % for 3 minutes
60 % for 11.5 minutes
30 % until setting is changed
Setting range 34 to 38 °C (93.2 to 100.4 °F)
>37 °C (98.6 °F), after confirmation
13 to 43 °C (55.4 to 109.4 °F)
2 temperature sensors for central and peripheral skin temperature,
graphical display of skin temperature
Measured Value Display
X-ray translucent, x-ray tray can be used for acquisition of x-ray
images
Setting range Off, 35 to 39 °C (95 to 102 °F), AUTO
Closed servo-humidification system with external water reservoir,
boiling and evaporation of Aquadest.