Control Buttons - Velleman DVM 205AM Manual Del Usuario

Ocultar thumbs Ver también para DVM 205AM:
Tabla de contenido

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

  • MEXICANO, página 19
Range
400.0kΩ
4.000MΩ
40.00MΩ
40.00nF
400.0nF
Capacitance
(auto-
4.000µF
ranging)
40.00µF
100.0µF
9.999Hz
99.99Hz
999.9Hz
Frequency
(auto-
9.999kHz
ranging)
99.99kHz
999.9kHz
9.999MHz
Duty cycle
0.1%~99.9%
2.0ms~
Period
20.0ms
2.0ms
Pulse width
~10.0ms
-20~+760°C
Temperature
-4~+1400°F
Diode test
0.3mA typical
Audible threshold: less than 50Ω
Audible
Test current: <0.3mA
Continuity
Overload protection on all ranges: 250Vdc/ac rms

5. Control Buttons

a) Hz / % Button (fig. p.2 #2)
This button allows you to choose between frequency and the duty cycle in a frequency range.
Press the HZ/% button to measure the frequency or the duty cycle while measuring voltage or current.
For the voltage/current requirements and frequency range, see the following table (for reference only):
Range (AC/DC)
4V
40V, 400V
1000V/750V
400mA
10A
Press the HZ/% button again to return to regular voltage or current measurement.
b) Range Button (fig. p.2 #3)
When the meter is switched on, it automatically goes into auto ranging ("AUTO" indication on the
display). This automatically selects the best range for the measurements being made and is generally the
best mode for most measurements. For measurements situations requiring a manually selected range,
perform the following steps:
1. Press the RANGE button; the "AUTO" indication on the display will disappear.
2. Press the RANGE button repeatedly to skip through the available ranges until the required one is
selected.
3. To return to auto ranging, press and hold the RANGE button for more than 2 seconds.
c) Relative Button (fig. p.2 #4)
The relative measurement feature allows you to make measurements relative to a stored reference
value. A reference voltage, current etc. can be stored and measurements made in comparison to that
value. The displayed value is the difference between the reference value and the measured value.
1. Perform any measurement as described in the operating instructions.
05.01.2010
DVM205AM
Resolution Accuracy
100 Ω
1kΩ
10kΩ
±2.0% rdg ±3 digits
10pF
±5.0% rdg ±7 digits
0.1nF
1nF
±3.0% rdg ±5 digits
10nF
0.1µF
±5.0% rdg ±5 digits
0.001Hz
±1.5% rdg ±5 digits
0.01Hz
0.1Hz
1Hz
±1.2% rdg ±3 digits
10Hz
100Hz
1kHz
±1.5% rdg ±4 digits
0.1%
±1.2% rdg ±2 digits
0.1ms
±3% rdg ±10 digits
0.1ms
±3% rdg ±10 digits
1°C
±3% rdg ±3 digits
(meter only ; probe
1°F
accuracy not included)
1mV
±10% rdg ±5 digits
Sensitivity
≥1.5V rms
≥6.5V rms
≥12V rms
≥420V rms
≥45mA rms
≥4A rms
5
other
Input protection: 250Vdc/ac rms
Sensitivity:
<0.5Vrms while ≤1MHz
>3Vrms while >1MHz
Overload protection: 250Vdc/ac
rms
Pulse width: >100µs, 100ms
Frequency width: 5Hz-150kHz
Sensitivity: <0.5Vrms
Overload protection: 250Vac/dc
rms
Overload protection: 250Vac/dc
rms
Overload protection: 250Vac/dc
rms
Sensor: K-type thermocouple
Open circuit voltage: 1.5Vdc
typical
Overload protection: 250Vac/dc
rms
Frequency width
5Hz~10kHz
5Hz~20kHz
5Hz~200kHz
50Hz~1kHz
5Hz~5kHz
5Hz~1kHz
©Velleman nv

Publicidad

Tabla de contenido
loading

Tabla de contenido