ALLURIS SMS-200E Instrucciones De Servicio página 18

Tabla de contenido

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

Flash tube emits a weak blue light and/or does not produce any noise:
The flash tube is used up. Please replace it (see descriptions in "Replacing the Flash Tube").
The battery life is shortened:
After approx. more than 300 recharging cycles the battery will be worn out and should therefore be
replaced (see "Replacing the Battery").
Flash sequence is uneven:
Please consult an authorized dealer or service specialist.
Attention: Do not open the instrument. Some components may still keep voltage also after
disconnection of the stroboscope from its internal or external power sources.
Warranty
We grant a 24 month limited warranty period from the date of purchase. Consumption material and parts, normal
wear and tear as well as damages caused by improper use are excluded from this warranty.