3M ESPE Elipar FreeLight Instrucciones De Manejo página 3

Ocultar thumbs Ver también para Elipar FreeLight:
Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

8. Should you have any reason to suspect the safety of
the unit to be compromised, the unit must be taken
out of operation and labeled accordingly to prevent
third parties from inadvertently using a possibly
defective unit. Safety may be compromised, e.g. if
the unit malfunctions or is noticeably damaged.
9. Solvents, flammable liquids, and powerful heat
sources must not be kept in the vicinity of the unit
as exposure to one of these factors may damage
the plastic housing of the unit.
10. Do not allow cleaning agents to enter the unit as
this may short-circuit the unit or cause potentially
dangerous malfunction.
11. Only factory-authorized technicians are authorized
to open the housing of the unit and carry out any
repairs.
12. Elipar FreeLight must not be used in patients, or by
users, with heart pacemaker implants who have
been advised to be cautious regards their exposure
to small electrical devices.
13. Do not use Elipar FreeLight in patients with a
history of photobiological reactions (including
Power cable
2
individuals with urticaria solaris or erythropoetic
protoporphyria) or who are currently on photo-
sensitizing medication (including 8-methoxypsora-
len or dimethylchlorotetracycline).
14. Individuals with a history of cataract surgery may
be particularly sensitive to the exposure to light
and should be discouraged from Elipar FreeLight
treatment unless adequate safety measures, such
as the use of protective goggles to remove blue
light, are undertaken.
15. Individuals with a history of retinal disease should
seek advice from their ophthalmologist prior to
operating the unit. In operating the Elipar FreeLight
unit, this group of individuals must take extreme
care and comply with any and all safety precautions
(including the use of suitable light-filtering safety
goggles).
16. Prior to each use of the device ensure that the
emitted light intensity is sufficient to safely
guarantee polymerization. Use the built-in light
testing area for this purpose.
Light guide
Handpiece
Charger

Publicidad

Tabla de contenido
loading

Tabla de contenido