8
The cycle can be launched if the door is closed.
The door is locked by the safety device upon the START command and remains blocked
until the cycle is completed.
The door cannot be opened during the cycle.
The cycle is launched from the beginning when START is pressed.
The STOP command interrupts the cycle, resets the functions performed and unlocks the
door.
When there is no disinfectant or oil in the tanks, START is inhibited and this is indicated on
the display. When this signal is given during the cycle, the latter is not interrupted.
When there is no air, START is inhibited and the cycle is interrupted, thereby resetting the
selections.
Safety during the cycle is controlled by software that displays the errors and anomalies
during operation and stores them on the SD card.
SAFETY DURING OPERATION
ENGLISH
27