CAUTION: To avoid damage to the communication
module internal electronics, do not allow solution
to soak into the electronic contacts area or in
any openings and seams. Use only a dry lint-free
cloth to clean the electronic contacts and the area
surrounding normally enclosed when installed in
the pump.
The following steps may be used to clean and disinfect
the communication module:
1. Clean the communication module using a mild
detergent soap solution to remove residuals or
contaminated material. Apply solution to a soft, lint-
free cloth and then wipe the exposed surfaces of the
communication module when installed in the pump.
Do not allow the solution to soak into the pump or the
communication module.
2. Disinfect the communication module by applying
a disinfecting solution (listed below) according to the
disinfecting product label instructions. If using a liquid
or spray, apply solution to a soft, lint-free cloth and
then wipe the exposed surfaces of the communication
module when installed in the pump. Follow the
disinfectant manufacturer's recommendations for
disinfectant contact times. Do not allow the solution to
soak into the pump or the communication module.
16
Acceptable disinfecting solutions for the CADD®-Solis
Communication Module are listed below.
Product
Manufacurer
(EPA
Registration
Number)
CaviWipes™ Metrex
(46781-8)
Sani-Cloth®
PDI
Super
(9480-4)
Sani-Cloth®
PDI
Bleach
(9480-8)
3. Allow the communication module to dry
completely before use.
Active
Contact/
Ingredient(s)
Kill Time
17.2%
3
Isopropanol
minutes
Dimethyl
2
Benzyl
minutes
Ammonium
Chloride,
Dimethyl
Ethyl Benzyl
Ammonium
Chloride
0.60%
4
Sodium
minutes
Hypochlorite